Skip to content

Swissmedic’s Focus Campaign: Key Outcomes on PMS Compliance


2 mins


Swissmedic conducted a focus campaign to assess compliance with the post-market surveillance (PMS) requirements of the Medical Devices Ordinance (MedDO) for higher-risk legacy devices. The campaign’s goal was to ensure manufacturers are adequately monitoring their medical devices after-market launch to detect potential issues early.

Key Findings:

    • Sample and Scope: The campaign reviewed 30 legacy devices in risk classes IIa and higher, with 28 of these products from foreign manufacturers.
    • Non-Conformities: Non-conformities were identified in 20 out of 30 cases, totaling 85 non-conformities across various PMS requirements.
    • PMS Plan and Safety Reports:
      • 11 products lacked a compliant PMS plan.
      • 8 products had safety reports not prepared according to the PMS plan.
      • 9 products had incomplete or missing information in the safety report.

Conclusions:

    • The MedDO requirements for PMS have been in effect since May 26, 2021. The campaign revealed that not all manufacturers are fully implementing these requirements.
    • Swissmedic will follow up on the rectification of non-conformities and may impose additional measures to ensure compliance.
    • Manufacturers are urged to review and enhance their PMS processes and documentation to meet MedDO requirements, ensuring patient health and safety.

____________________________________________

Questions about this new Swissmedic’s Focus Campaign and PMS Compliance? Contact us.

Related Resources: